CLEARANCE: The shortest distance between two conductive parts, or between a conductive part and the bounding surface of the
equipment measured through air.
CREEPAGE DISTANCE: The shortest path between two conductive parts or between a conductive part and the bounding surface of the
equipment, measured along the surface of the insulation.
Note: Bounding surface is the outer surface of the electrical enclosure, consider as though metal foil were pressed into contact with accessible surface of insulating material.
Requirement
Table 1 describes the differences between IEC 950 and IEC 601-1.
|
Maximum Input Voltage
|
I.T.E.
|
Medical Equipment
|
|
Basic
|
Reinforce
|
Basic
|
Reinforce
|
|
125V
|
Air Clearance
|
1.3mm
|
2.6mm
|
1.6mm
|
3.2mm
|
|
Creepage Distance
|
1.5mm
|
3.0mm
|
3.0mm
|
6.0mm
|
|
250V
|
Air Clearance
|
2.0mm
|
4.0mm
|
2.5mm
|
5.0mm
|
|
Creepage Distance
|
2.5mm
|
5.0mm
|
4.0mm
|
8.0mm
|
For I.T.E. and medical equipment, the spacing requirement mention in table 1 are being use to provide a method of insulation as illustrated above.
Class 2
|
Voltage Differences (Potential Difference in Vrms)
|
With Opening
|
Without Opening
|
|
50 Vrms or less
|
Air Clearance
|
1.6mm
|
1.6mm
|
|
Creepage Distance
|
1.6mm
|
1.6mm
|
|
51-150 Vrms
|
Air Clearance
|
3.2mm
|
1.6mm
|
|
Creepage Distance
|
6.4mm
|
1.6mm
|
|
151-250
|
Air Clearance
|
6.4mm
|
4.8mm
|
|
Creepage Distance
|
9.5mm
|
4.8mm
|
Equipment
Spacing requirement for UL1310 equipment is different from I.T.E. and medical equipment.
The spacing mention in table 2 are between live parts of opposite polarity and between live and dead metal parts.
Note: please refer to your safety engineer if the enclosure is made of metal.
|